  "geometry": "wide-inlet CAD, unchanged",
  "maximum_commanded_flow_mm3_s": 26.4,
  "viscosity_Pa_s": [
    0.5,
    1.0
  ],
  "pressure_budget_Pa": 6500,
  "worst_computed_pressure_Pa": 6478.59692771179,
  "operating_rule": "Use the unchanged selected CAD within the declared tolerance, feed-tube and meniscus bounds, with commanded flow no greater than the computed 26.4 mm3/s bound. The former 30 mm3/s endpoint is rejected for this combined system envelope. Fixed-pressure pulse dosing is qualified only for the stated nominal de-aired case; changed viscosity requires recalculating pressure or pulse duration."
}